Sulmaz Zahedi, Miles Hudson, Xin Jin, Richard Justin Garling, Jacob Gluski, Caden Nowak, Neena I. Marupudi, Paul Begeman, and Carolyn A. Harris

OBJECTIVE

This investigation is aimed at gaining a better understanding of the factors that lead to mechanical failure of shunts used for the treatment of hydrocephalus, including shunt catheter-valve disconnection and shunt catheter fracture.

开云体育世界杯赔率

To determine the root cause of mechanical failure, the authors created a benchtop mechanical model to mimic mechanical stressors on a shunt system. To test shunt fracture, cyclical loading on the catheter-valve connection site was tested with the shunt catheter held perpendicular to the valve. Standard methods were used to secure the catheter and valves with Nurolon. These commercial systems were compared to integrated catheters and valves (manufactured as one unit). To test complete separation/disconnection of the shunt catheter and valve, a parallel displacement test was conducted using both Nurolon and silk sutures. Finally, the stiffness of the catheters was assessed. All mechanical investigations were conducted on shunts from two major shunt companies, assigned as either company A or company B.

RESULTS

循环加载实验发现分流摇来摇去m company B fractured after a mean of 4936 ± 1725 cycles (95% CI 2990–6890 cycles), while those of company A had not failed after 8000 cycles. The study of parallel displacement indicated complete disconnection of company B’s shunt catheter-valve combination using Nurolon sutures after being stretched an average 32 ± 5.68 mm (95% CI 25.6–38.4 mm), whereas company A’s did not separate using either silk or Nurolon sutures. During the stiffness experiments, the catheters of company B had statistically significantly higher stiffness of 13.23 ± 0.15 N compared to those of company A, with 6.16 ± 0.29 N (p < 0.001).

CONCLUSIONS

Mechanical shunt failure from shunt catheter-valve disconnection or fracture is a significant cause of shunt failure. This study demonstrates, for the first time, a correlation between shunt catheters that are less mechanically stiff and those that are less likely to disconnect from the valve when outstretched and are also less likely to tear when held at an angle from the valve outlet. The authors propose an intervention to the standard of care wherein less stiff catheters are trialed to reduce disconnection.

Restricted access

Prashant Hariharan, Jacob Gluski, Jeffrey Sondheimer, Alexandra Petroj, Andrew Jea, William E. Whitehead, Marc R. Del Bigio, Neena I. Marupudi, James P. McAllister II, David D. Limbrick Jr., Brandon G. Rocque, and Carolyn A. Harris

OBJECTIVE

The aim of this study was to explore how clinical factors, including the number of lifetime revision surgeries and the duration of implantation, affect the degree of obstruction and failure rates of ventricular catheters (VCs) used to manage hydrocephalus.

开云体育世界杯赔率

A total of 343 VCs and their associated clinical data, including patient demographics, medical history, and surgical details, were collected from 5 centers and used for this analysis. Each VC was classified by the degree of obstruction after macroscopic analysis. Univariate, multivariate, and binned analyses were conducted to test for associations between clinical data and degree of VC obstruction.

RESULTS

VCs from patients with 0 to 2 lifetime revisions had a larger proportion of VC holes obstructed than VCs from patients with 10 or more revisions (p = 0.0484). VCs implanted for less than 3 months had fewer obstructed holes with protruding tissue aggregates than VCs implanted for 13 months or longer (p = 0.0225). Neither duration of implantation nor the number of lifetime revisions was a significant predictor of the degree of VC obstruction in the regression models. In the multinomial regression model, contact of the VCs with the ventricular wall robustly predicted the overall obstruction status of a VC (p = 0.005). In the mixed-effects model, the age of the patient at their first surgery emerged as a significant predictor of obstruction by protruding tissue aggregates (p = 0.002). VCs implanted through the parietal entry site were associated with more holes with nonobstructive growth and fewer empty holes than VCs implanted via other approaches (p = 0.001).

CONCLUSIONS

The number of lifetime revisions and duration of implantation are correlated with the degree of VC obstruction but do not predict it. Contact of the VC with the ventricular wall and the age of the patient at their first surgery are predictors of the degree of VC obstruction, while the entry site of the VC correlates with it.