Thomas A. Pieters, Christopher R. Conner, and Nitin Tandon

Object

Precise localization of subdural electrodes (SDEs) is essential for the interpretation of data from intracranial electrocorticography recordings. Blood and fluid accumulation underneath the craniotomy flap leads to a nonlinear deformation of the brain surface and of the SDE array on postoperative CT scans and adversely impacts the accurate localization of electrodes located underneath the craniotomy. Older methods that localize electrodes based on their identification on a postimplantation CT scan with coregistration to a preimplantation MR image can result in significant problems with accuracy of the electrode localization. The authors report 3 novel methods that rely on the creation of a set of 3D mesh models to depict the pial surface and a smoothed pial envelope. Two of these new methods are designed to localize electrodes, and they are compared with 6 methods currently in use to determine their relative accuracy and reliability.

开云体育世界杯赔率

的first method involves manually localizing each electrode using digital photographs obtained at surgery. This is highly accurate, but requires time intensive, operator-dependent input. The second uses 4 electrodes localized manually in conjunction with an automated, recursive partitioning technique to localize the entire electrode array. The authors evaluated the accuracy of previously published methods by applying the methods to their data and comparing them against the photograph-based localization. Finally, the authors further enhanced the usability of these methods by using automatic parcellation techniques to assign anatomical labels to individual electrodes as well as by generating an inflated cortical surface model while still preserving electrode locations relative to the cortical anatomy.

Results

的recursive grid partitioning had the least error compared with older methods (672 electrodes, 6.4-mm maximum electrode error, 2.0-mm mean error, p < 10−18)。的maximum errors derived using prior methods of localization ranged from 8.2 to 11.7 mm for an individual electrode, with mean errors ranging between 2.9 and 4.1 mm depending on the method used. The authors also noted a larger error in all methods that used CT scans alone to localize electrodes compared with those that used both postoperative CT and postoperative MRI. The large mean errors reported with these methods are liable to affect intermodal data comparisons (for example, with functional mapping techniques) and may impact surgical decision making.

Conclusions

的authors have presented several aspects of using new techniques to visualize electrodes implanted for localizing epilepsy. The ability to use automated labeling schemas to denote which gyrus a particular electrode overlies is potentially of great utility in planning resections and in corroborating the results of extraoperative stimulation mapping. Dilation of the pial mesh model provides, for the first time, a sense of the cortical surface not sampled by the electrode, and the potential roles this “electrophysiologically hidden” cortex may play in both eloquent function and seizure onset.

Restricted access

Gabrielle Santangelo, Thomas A. Pieters, Muhammad I. Jalal, Redi Rahmani, Howard J. Silberstein, and Jonathan J. Stone

OBJECTIVE

While CSF diversion is highly effective at treating hydrocephalus, shunting is unfortunately associated with a very high revision rate. Studies have demonstrated that proximal catheter obstruction is a major cause of failure. A novel proximal access device was developed, and pilot testing was performed in a sheep model of hydrocephalus.

开云体育世界杯赔率

Hydrocephalus was induced in 8 sheep using a cisternal injection of 4 ml of 25% kaolin, and the sheep were randomized to either a standard ventricular catheter or a novel intraparenchymal stent (IPS). Both groups received identical valves and distal catheters. The novel device included a 3D-printed stainless steel port and a 6 × 40–mm covered peripheral vascular stent. Animals were euthanized for signs of hydrocephalus or at a time point of 2 months. MRI was performed to determine ventricular size. Time to failure and Evans indices were compared using the Wilcoxon rank-sum test.

RESULTS

All 4 experimental devices were placed without difficulty into the right lateral ventricle. There was a trend toward longer survival in the experimental group (40 vs 26 days, p = 0.24). Within the IPS group, 3 of the 4 sheep did not experience clinical symptoms of shunt failure and had an average of 37% decrease in Evans index. While 3 of 4 traditional proximal catheters demonstrated debris within the inlet holes, there was no obstructive material found in the IPSs.

CONCLUSIONS

An IPS was successfully used to treat hydrocephalus in a sheep model. While statistical significance was not achieved, there were clear benefits to using a stent, including a decreased clog rate and the ability to perform a percutaneous revision. Further testing is needed to ensure efficacy and safety prior to human application.