Image-guided frameless stereotactic techniques provide an alternative to traditional head-frame fixation in the performance of fine-needle biopsies. However, these techniques still require rigid head fixation, usually in the form of a head holder. The authors report on a series of fine-needle biopsies and brain abscess aspirations in which a frameless technique was used with a patient's head supported on a horseshoe headholder. To validate this technique, they performed an in vitro accuracy study.
Forty-eight patients underwent fine-needle biopsy of intracranial lesions that ranged in size from 0.9 to more than 107.7 ml; a fiducial-less, frameless, image-guided technique was used without rigid head fixation. In 1 of the 48 patients a cerebral abscess was drained. The accuracy study was performed with a skull phantom that was imaged with a CT scanner and tracked with a registration mask containing light-emitting diodes. The objective was a skin fiducial marker with a 4-mm circular target to accommodate the 2.5-mm biopsy needle. A series of 50 trials was conducted.
Diagnostic tissue was obtained on the first attempt in 47 of 48 brain biopsy cases. In 2 cases small hemorrhages at the biopsy site were noted as a complication on the postoperative CT scan. One of these hemorrhages resulted in hand and arm weakness. The accuracy study demonstrated a 98% success rate of the biopsy needle passing through the 4-mm circular target using the registration mask as the registration and tracking device. This demonstrates a ± 0.75-mm tolerance on the targeting method.
精度研究证明的能力mask to actively track the target and allow navigation to a 4-mm-diameter circular target with a 98% success rate. The frameless, pinless, fiducial-less technique described herein will likely be another safe, fast alternative to frame-based stereotactic techniques for fine-needle biopsy that avoids the potential morbidity of rigid head-pin fixation. Furthermore, it should lend itself to other image-guided applications such as the placement of ventricular catheters for shunting or Ommaya reservoirs.
Please include this information when citing this paper: published online August 20, 2010; DOI: 10.3171/2010.7.JNS091493.
BarnettGH,,KormosDW,,SteinerCP,&WeisenbergerJ:Use of a frameless, armless stereotactic wand for brain tumor localization with two-dimensional and three-dimensional neuroimaging.开云体育app官方网站下载入口33:674–678,1993
BarnettGH,,MillerDW,&WeisenbergerJ:Frameless stereotaxy with scalp-applied fiducial markers for brain biopsy procedures: experience in 218 cases.J Neurosurg91:569–576,1999
DorwardNL,,AlbertiO,,PalmerJD,,KitchenND,&ThomasDG:Accuracy of true frameless stereotaxy: in vivo measurement and laboratory phantom studies. Technical note.J Neurosurg90:160–168,1999
GermanoIM,&QueenanJV:Clinical experience with intracranial brain needle biopsy using frameless surgical navigation.Comput Aided Surg3:33–39,1998
HelmPA,&EckelTS:Accuracy of registration methods in frameless stereotaxis.Comput Aided Surg3:51–56,1998
LeuthardtEC,,FoxD,,OjemannGA,,DaceyRG,,GrubbRL,&RichKM,et al.:Frameless stereotaxy without rigid pin fixation during awake craniotomies.Stereotact Funct Neurosurg79:256–261,2002
ManganoFT,,LimbrickDDJr,,LeonardJR,,ParkTS,&SmythMD:Simultaneous image-guided and endoscopic navigation without rigid cranial fixation: application in infants: technical case report.开云体育app官方网站下载入口58:4 Suppl 2ONS-E377,2006
PaleologosTS,,DorwardNL,,WadleyJP,&ThomasDG:Clinical validation of true frameless stereotactic biopsy: analysis of the first 125 consecutive cases.开云体育app官方网站下载入口49:830–837,2001
Reavey-CantwellJF,,BovaFJ,&PincusDW:Frameless, pinless stereotactic neurosurgery in children.J Neurosurg104:6 Suppl392–395,2006
ReichterRE,,ParkHM,&HallD:Skull defects due to Mayfield head stabilizer.Clin Nucl Med8:553,1983